Back to Top

Hi, Guest!

  LOKASI :  Kota Administrasi Jakarta Selatan

Bergabung Selama :

BAGIKAN :   

Bagikan :

Amprobe BAT-500 Battery Impedance Tester up to 40V

Update Terakhir
:
01 / 12 / 2019
Min. Pembelian
:
1 Unit
Dilihat Sebanyak
:
9 kali

Harga

CALL
Bagikan
:

Perhatian !

Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.

Detail Amprobe BAT-500 Battery Impedance Tester Up To 40V

DESCRIPTION :

Amprobe BAT-500 Battery Tester quickly and accurately tests batteries without taking the batteries off-line. This Battery Tester features a comparator function with pass/ fail indication, and speed testing of batteries up to 500Ah. Amprobe BAT-500 Battery Tester, compact and light weight, simultaneously measures battery resistance and voltage up to 40V.

FEATURES :

Tests battery Impedance to calculate its true life Works with most batteries, compact storage batteries, alkaline and lead-acid up to 500Ah Resistance measurement up to 40 ohms, and voltage up to 40V Tests are performed without shutting down a battery Simultaneously measure battery resistance and voltage Comparator function to quickly access conditions as pass, warning, or fail 99 memory locations to store results with read function Compact and lightweight RS232 PC interface to download testing results to a PC for trending/ life cycle analysis Instruction manual, batteries, test probe, software, RS232 cable Features BAT-500 Maximum battery capacity Up to 500 amp hours Battery type Alkaline, nickel-metal hydride ( NiMH) , nickel-cadmium ( NiCd) , lithium-ion ( Li-ion) and lead-acid batteries DC Voltage 4.000 V, 40.00 V Impedance 40 mO, 400 mO, 4 O, 40 O Data capture memory 500 data sets Data logging memory 9600 data sets PC Download

• Specifications :

Battery capacity: 0 to 500AH
Resolution: 10µ O, 100µ O, 1mO, 10mO
Accuracy: ± ( 1% reading ± 10digits)
Open circuit voltage: 3.5Vpp max
Measurement current: 25mA approx., 2.5mA approx., 250µ A approx., 25µ A approx
Measuring current frequency: 1KHz ± 10%
Voltage measurement range: 4V, 40V
Resolution: 1mV, 10mV Accuracy: ± ( 0.1% rdg ± 6 digits)

Tampilkan Lebih Banyak